530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-42892
- Event Type
- Malfunction
- Date Received
- October 25, 2014
- Date of Event
- September 26, 2014
- Report Date
- September 26, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). CURRENTLY IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, A SCRATCHED RESERVOIR TUBE WINDOW AND A STAINED END CAP STICKER. NO MOISTURE DAMAGE WAS NOTED ON THE ELECTRONICS, MOTOR, VIBRATOR MOTOR OR BATTERY TUBE ASSEMBLY.
CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP AND WAS UNABLE TO CLEAR IT. CUSTOMER STATED THAT THE INSULIN PUMP MAY HAVE BEEN EXPOSED TO MOISTURE FROM SWEAT. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 65 MG/DL AND HAS TREATED WITH GLUCOSE TABS AND BREAKFAST. NO FURTHER INFO REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 680397 | 530G INSULIN PUMP | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |