FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4213451 · Received October 30, 2014

Report

Report Number
2531779-2014-30872
Event Type
Malfunction
Date Received
October 30, 2014
Report Date
October 14, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/08/2014 WITH THE FOLLOWING FINDINGS: THE BATTERY CAP WAS NOT RETURNED WITH THE PUMP FOR INVESTIGATION. A TEST BATTERY CAP WAS USED TO COMPLETE TESTING. INVESTIGATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED AT THE OPENING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE USER HAD DIFFICULTY REMOVING THE BATTERY CAP FROM THE PUMP FOR A BATTERY CHANGE. THE REPORTER NOTED THAT THE BATTERY CAP EVENTUALLY WAS REMOVED WITH USE OF PIERS. THE REPORTER STATED THAT THE USE THEN INSERTED A NEW BATTERY AND PUT A NEW BATTERY CAP ON THE PUMP, BUT THE NEW CAP DID NOT ATTACH PROPERLY, AS IF THERE WAS AN ISSUE WITH THE BATTERY COMPARTMENT THREADS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.