ANIMAS VIBE
Report
- Report Number
- 2531779-2014-30872
- Event Type
- Malfunction
- Date Received
- October 30, 2014
- Report Date
- October 14, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/08/2014 WITH THE FOLLOWING FINDINGS: THE BATTERY CAP WAS NOT RETURNED WITH THE PUMP FOR INVESTIGATION. A TEST BATTERY CAP WAS USED TO COMPLETE TESTING. INVESTIGATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED AT THE OPENING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE USER HAD DIFFICULTY REMOVING THE BATTERY CAP FROM THE PUMP FOR A BATTERY CHANGE. THE REPORTER NOTED THAT THE BATTERY CAP EVENTUALLY WAS REMOVED WITH USE OF PIERS. THE REPORTER STATED THAT THE USE THEN INSERTED A NEW BATTERY AND PUT A NEW BATTERY CAP ON THE PUMP, BUT THE NEW CAP DID NOT ATTACH PROPERLY, AS IF THERE WAS AN ISSUE WITH THE BATTERY COMPARTMENT THREADS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 695447 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |