530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-43263
- Event Type
- Malfunction
- Date Received
- October 25, 2014
- Date of Event
- September 26, 2014
- Report Date
- September 26, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
THE INSULIN PUMP PASSED DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, COMPROMISED FORCE SENSOR, AND NO DELIVERY TEST. THERE WAS NO MOTOR ERROR ALARM NOTED. THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CASE AT THE DISPLAY WINDOW CORNER, MINOR SCRATCHED LCD WINDOW, AND SCRATCHED RESERVOIR TUBE WINDOW.
CUSTOMER REPORTES TO HAVE RECEIVED TWO MOTOR ERROR ALARMS. CUSTOMER'S BLOOD GLUCOSE WAS 258 MG/DL. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. NO FURTHER INFORMATION PROVIDED.