FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4213440 · Received October 25, 2014

Report

Report Number
2032227-2014-42901
Event Type
Malfunction
Date Received
October 25, 2014
Date of Event
September 26, 2014
Report Date
September 26, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE, DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM OR MOISTURE DAMAGE ON ELECTRONIC ASSEMBLY OR MOTOR NOTED. INSULIN PUMP HAD A CRACKED CASE ON DISPLAY WINDOW CORNER, MINOR SCRATCHES ON LCD WINDOW AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM. THE BLOOD GLUCOSE READING IS 83 MG/DL. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680676 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 51 YR