FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4213440
·
Received October 25, 2014
Report
- Report Number
- 2032227-2014-42901
- Event Type
- Malfunction
- Date Received
- October 25, 2014
- Date of Event
- September 26, 2014
- Report Date
- September 26, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE, DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM OR MOISTURE DAMAGE ON ELECTRONIC ASSEMBLY OR MOTOR NOTED. INSULIN PUMP HAD A CRACKED CASE ON DISPLAY WINDOW CORNER, MINOR SCRATCHES ON LCD WINDOW AND CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
CUSTOMER REPORTED A BUTTON ERROR ALARM. THE BLOOD GLUCOSE READING IS 83 MG/DL. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 680676 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |