FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIIN INFUSION PUMP

MDR report key: 4213438 · Received October 13, 2014

Report

Report Number
2032227-2014-36919
Event Type
Malfunction
Date Received
October 13, 2014
Date of Event
June 14, 2014
Report Date
June 18, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH NO BACK LIGHT DUE TO FAULTY PANEL ON THE LCD DISPLAY. NO COSMETIC DAMAGE NOTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP MAY HAVE A BACKLIGHT ANOMALY. CUSTOMER STATED THEY ARE UNABLE TO VIEW PROGRAMMING INFO DUE TO THE DARKNESS OF THE DISPLAY SCREEN. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LED TO THE ALARM OR IF THE DEVICE HAD BEEN DROPPED. ADVISED CUSTOMER TO DISCONTINUE USE OF RESERVOIR AND REVERT TO BACK UP PLAN. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 143 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
647417 PARADIGM REAL-TIME REVEL INSULIIN INFUSION PUMP OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 31 YR