FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4213436 · Received October 25, 2014

Report

Report Number
2032227-2014-42903
Event Type
Malfunction
Date Received
October 25, 2014
Date of Event
September 26, 2014
Report Date
September 26, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED. NO MOISTURE DAMAGE WAS FOUND INSIDE OF THE INSULIN PUMP. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP, A CRACKED CASE AT THE RESERVOIR TUBE WINDOW CORNER AND A BROKEN BELT CLIP SLOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE BLOOD GLUCOSE READING WAS 112 MG/DL. THE CUSTOMER STATED THAT HE HAD BEEN CAUGHT IN A NASTY RAIN STORM AND EVERYTHING TOO WET. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680412 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 40 YR