FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4213425 · Received October 30, 2014

Report

Report Number
2531779-2014-30871
Event Type
Malfunction
Date Received
October 30, 2014
Report Date
October 15, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/22/2014 WITH THE FOLLOWING FINDINGS: COMPLETED INVESTIGATION USING TEST BATTERY CAP. RETURNED BATTERY CAP IS CORRODED AND WILL NOT MAKE CONTACT. BATTERY COMPARTMENT IS CRACKED AT OPENING, DOWN TO RUBBER BUMPER. EVIDENCE OF MOISTURE INSIDE BATTERY COMPARTMENT..LEAK TEST FAILED DUE TO BATTERY COMPARTMENT LEAK. DISPLAY IS DIM/FADED AND LETTERS ARE RED. AUDIO BOLUS BUTTON IS INTERMITTENTLY WORKING WITH NO PHYSICAL DAMAGE. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(4).

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CASE DAMAGE W/MOISTURE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
695776 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1