SYMBIQ DCHANNEL 3.01
Report
- Report Number
- 9615050-2014-05740
- Date Received
- October 24, 2014
- Date of Event
- January 1, 2014
- Report Date
- October 6, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0070-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE TOUCHSCREEN WOULD NOT RESPOND. THIS WAS DUE TO THE TOUCHSCREEN BEING OUT OF CALIBRATION. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN WAS NON RESPONSIVE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN DID NOT RESPOND. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 679025 | SYMBIQ DCHANNEL 3.01 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |