FDA Adverse Event Malfunction Summary report: N

PRISM HIV O PLUS

MDR report key: 4213355 · Received October 30, 2014

Report

Report Number
1415939-2014-00228
Event Type
Malfunction
Date Received
October 30, 2014
Report Date
October 7, 2014
Manufacturer
ABBOTT LABORATORIES
Product Code
MZF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. (B)(4)..

Additional Manufacturer Narrative · 1

FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, A REVIEW OF LABELING AND SPECIFICITY TESTING. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER'S ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. SPECIFICITY TESTING ON A REPRESENTATIVE LOT MET ALL SPECIFICATIONS. COMPLAINT DATA REVIEW BY LOT AND DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED SINCE THE SUSPECT MEDICAL DEVICE LOT IS UNKNOWN. HOWEVER, REVIEW OF THE PRISM METRICS FIELD DATA FROM JANUARY - AUGUST 2014 INDICATES THAT THE PRODUCT IS MEETING LABELING CLAIMS FOR CLINICAL SPECIFICITY. BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS' COMPLAINT INVESTIGATION, THE ASSAY PERFORMED AS INTENDED AND NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

THE CUSTOMER OBSERVED (B)(6) RESULTS THAT DO NOT CONFIRM FOR BLOOD DONORS ON THE PRISM ANALYZER. DATA WAS PROVIDED SHOWING THEIR REPEAT (B)(6) RATE IS BELOW (B)(6), BUT HIGHER THAN (B)(6) DATA. (B)(6) DONORS ARE (B)(6). PER THE PRISM (B)(4) PACKAGE INSERT: REPEATEDLY REACTIVE RESULTS INDICATE THE PRESENCE OF (B)(6) BY THE CRITERIA OF THE ABBOTT PRISM (B)(4) ASSAY. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
696740 PRISM HIV O PLUS HIV-1 AND HIV-2 ANTIBODIES MZF ABBOTT LABORATORIES UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 LN 06A36-04, SN (B)(4)| PRISM ANALYZER:| PRISM ANALYZER:| LN 06A36-04, SN (B)(4)