FDA Adverse Event Malfunction Summary report: N

V60 VENTILATOR

MDR report key: 4213342 · Received October 24, 2014

Report

Report Number
2031642-2014-01291
Event Type
Malfunction
Date Received
October 24, 2014
Report Date
September 30, 2014
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
MNT
PMA / PMN Number
K082660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RESEATED THE DATA ACQUISITION TO MOTOR CONTROLLER CABLE.

Description of Event or Problem · 1

THE INTERNATIONAL CUSTOMER REPORTED THE VENTILATOR ALARMED AND DISPLAYED A VENT INOP DATA ACQUISITION PCB ADC FAILURE. THE CUSTOMER REPORTED THE UNIT WAS NOT IN USE ON A PT; THEREFORE, THERE WAS NO PT INVOLVEMENT OR HARM. A VENT INOP CONDITION DURING NORMAL VENTILATION OPERATION WILL RESULT IN A VISUAL AND AUDIBLE ALARM AND PRECLUDES CONTINUED SAFE OPERATION OF THE VENTILATOR. THE USER MUST PROVIDED ALTERNATIVE VENTILATION AND HAVE THE VENTILATOR SERVICED. THE MFR'S SERVICE ENGINEER REPORTED A CONNECTION FAILURE OF THE DATA ACQUISITION PCB TO MOTOR CONTROLLER PCB CABLE. THE SERVICE ENGINEER REPORTED THE CABLE WAS RECONNECTED TO CORRECT THE REPORTED PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680083 V60 VENTILATOR VENTILATOR, CONTINUOUS MNT RESPIRONICS CALIFORNIA, INC. V60 NA

Patients

Seq Age Sex Outcome Treatment
1