FDA Adverse Event Malfunction Summary report: N

STRYKER NAV3I(TM) PLATFORM

MDR report key: 4213311 · Received October 30, 2014

Report

Report Number
0001811755-2014-03842
Event Type
Malfunction
Date Received
October 30, 2014
Date of Event
October 2, 2014
Report Date
October 2, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HAW
PMA / PMN Number
K993239
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE REPORTED EVENT COULD BE DUPLICATED, AND INCORRECT SURFACE REGISTRATION WAS DETERMINED TO BE A POSSIBLE ROOT CAUSE.

Description of Event or Problem · 1

DURING A CRANIOTOMY FOR TUMOR PARIETAL RESECTION WITH LANGUAGE AND MOTOR MAPPING USING STRYKER NAV3I AND CRANIAL MAP SOFTWARE, THE SURGEON NOTICED AN INACCURACY OF THE SYSTEM. AFTER OPENING THE BONE FLAP TO ACCESS THE TUMOR LOCATION, THE SURGEON REPORTED NAVIGATION TO BE OFF APPROXIMATELY 2CM, BASED ON THE MRI. MASK REGISTRATION WITH RESCUE POINTS WERE USED ALONG WITH A LANDMARK CHECK FOR REGISTRATION. NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO CLINICALLY SIGNIFICANT DELAY WERE REPORTED WITH THIS EVENT. NAVIGATION WAS CANCELED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ULTRASOUND.

Description of Event or Problem · 1

DURING A CRANIOTOMY FOR TUMOR PARIETAL RESECTION WITH LANGUAGE AND MOTOR MAPPING USING STRYKER NAV3I AND CRANIAL MAP SOFTWARE, THE SURGEON NOTICED AN INACCURACY OF THE SYSTEM. AFTER OPENING THE BONE FLAP TO ACCESS THE TUMOR LOCATION, THE SURGEON REPORTED NAVIGATION TO BE OFF APPROXIMATELY 2CM, BASED ON THE MRI. MASK REGISTRATION WITH RESCUE POINTS WERE USED ALONG WITH A LANDMARK CHECK FOR REGISTRATION. NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO CLINICALLY SIGNIFICANT DELAY WERE REPORTED WITH THIS EVENT. NAVIGATION WAS CANCELED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ULTRASOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
697542 STRYKER NAV3I(TM) PLATFORM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1