530G INSULIN PUMP

MDR report key: 4213251 · Received October 13, 2014

Report

Report Number
2032227-2014-37264
Event Type
Malfunction
Date Received
October 13, 2014
Date of Event
September 13, 2014
Report Date
September 13, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH ALL BUTTONS RESPONDING PROPERLY. NO BUTTON ERROR ALARMS NOTED. HOWEVER, SLIGHTLY MOISTURE DAMAGE FOUND AT KEYPAD TRACES. THE DEVICE WAS RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, AND MINOR SCRATCHED LCD WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 290 MG/DL. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS, LEADING TO THE ALARM. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADD'L INFO IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
648704 530G INSULIN PUMP OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-551NAL

Patients

Seq Age Sex Outcome Treatment
1 44 YR