FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 4213250 · Received October 3, 2014

Report

Report Number
1824206-2014-02398
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
September 4, 2014
Report Date
September 4, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

HILL-ROM TECHNICAL SUPPORT FOUND THE SCALE BOARD TO BE INOPERABLE. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTIVE MAINTENANCE ON THEIR BEDS. THE ACCOUNT REPLACED THE SCALE BOARD TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED SCALE WAS NOT READING CORRECTLY. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620969 TOTALCARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1