FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 4213247
·
Received October 3, 2014
Report
- Report Number
- 1824206-2014-02390
- Event Type
- Malfunction
- Date Received
- October 3, 2014
- Date of Event
- September 8, 2014
- Report Date
- September 8, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- PMA / PMN Number
- K122473
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE SCALE NEEDED TO BE RECALIBRATED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN RECALIBRATED THE SCALE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED SCALE WAS NOT READING CORRECTLY. THE BED WAS LOCATED AT THE ACCOUNT IN THE WAREHOUSE. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 620982 | TOTALCARE BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |