FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 4213247 · Received October 3, 2014

Report

Report Number
1824206-2014-02390
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE SCALE NEEDED TO BE RECALIBRATED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN RECALIBRATED THE SCALE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED SCALE WAS NOT READING CORRECTLY. THE BED WAS LOCATED AT THE ACCOUNT IN THE WAREHOUSE. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620982 TOTALCARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1840

Patients

Seq Age Sex Outcome Treatment
1