FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 4213241
·
Received October 3, 2014
Report
- Report Number
- 1824206-2014-02391
- Event Type
- Malfunction
- Date Received
- October 3, 2014
- Date of Event
- September 8, 2014
- Report Date
- September 8, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- PMA / PMN Number
- K122473
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECH FOUND THE SCALE NEED TO BE RECALIBRATED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNK IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECH RECALIBRATED THE SCALE TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED SCALE WAS NOT READING CORRECTLY. THE BED WAS LOCATED AT THE ACCOUNT ON THE 3RD FLOOR C SICU 18.. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 620966 | TOTALCARE BED | A/C POWERED ADJUSTABLE HOSP BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |