FDA Adverse Event Malfunction Summary report: N

TERUMO ADVANCED PERFUSIN SYSTEM 1

MDR report key: 4213142 · Received August 4, 2014

Report

Report Number
1828100-2014-00676
Event Type
Malfunction
Date Received
August 4, 2014
Date of Event
July 11, 2014
Report Date
July 11, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DTQ
PMA / PMN Number
K022947
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS RELATED TO MDR #1828100-2014-00677. THE FIELD SERVICE REPRESENTATIVE (FSR) HOOKED UP THE PUMP HEADS ON ANOTHER SYSTEM-1 AND THE SOFTWARE UPDATE FAILED. THE FSR TOOK THE SMALL P/C CARD OUT OF ITS JACKET AND USED ANOTHER JACKET AND SOFTWARE UPDATE STILL FAILED. THE FSR COULD NOT DO THE NFC WITH THE NEW SOFTWARE CARD. THE FSR REPLACED THE ROLLER PUMP HEAD WITH A LOANER AND COMPLETED PREVENTATIVE MAINTENANCE. THE UNIT OPERATED TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. THE SUSPECT ROLLER PUMP WAS RETURNED TO THE MANUFACTURER FOR FURTHER INVESTIGATION. THE REPORTED COMPLAINT WAS CONFIRMED BY LABORATORY EVALUATION. THE QUALITY LABORATORY TECHNICIAN REMOVED OF THE ROLLER PUMP MODULE AND REPLACEMENT WITH A KNOWN GOOD MODULE RESOLVED THE ISSUE. CLOSER INSPECTION FOUND THE GENERIC BOARD WITHIN THE MODULE WOULD NOT ALLOW THE NEW SOFTWARE TO BE DOWNLOADED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING NOTICE OF FIELD CORRECTION (NFC) OF THE DEVICE, THE ROLLER PUMP HEAD WAS DEAD AFTER DOZENS OF ATTEMPTS TO LOAD THE 1.40 SOFTWARE. THE PUMP BOOTED UP BEFORE THE FSR TRIED TO PERFORM THE NFC. SINCE THE EVENT OCCURRED DURING NFC, THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
455618 TERUMO ADVANCED PERFUSIN SYSTEM 1 APS 1 (ROLLER PUMP) DTQ TERUMO CARDIOVASCULAR SYSTEMS CORP. 816571

Patients

Seq Age Sex Outcome Treatment
1