FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4213127 · Received October 30, 2014

Report

Report Number
2531779-2014-30862
Event Type
Injury
Date Received
October 30, 2014
Date of Event
October 3, 2014
Report Date
October 13, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN REQUESTED TO BE RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT ON (B)(6) 2014, THE PATIENT EXPERIENCED A BLOOD GLUCOSE (BG) OVER 500 MG/DL WITH A RACING HEART, RAPID, DEEP BREATHING/SHORTNESS OF BREATH AND/OR CHEST PAIN ASSOCIATED WITH AN INTERMITTENT POWER ISSUE. IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED, DISCONTINUED PUMP THERAPY, AND WAS TREATED BY THEIR HEALTHCARE PROVIDER WITH INSULIN VIA INJECTION AND IV FLUIDS. REPORTEDLY, THE PATIENT HAD A SMALL CARDIAC BLOCKAGE. DURING TROUBLESHOOTING WITH CUSTOMER TECHNICAL SUPPORT (CTS), IT WAS REVEALED THAT THE PATIENT HAD EXPERIENCED INTERMITTENT POWER DUE TO INCORRECTLY TIGHTENING THE BATTERY CAP ONTO THE PUMP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT ALLEGEDLY EXPERIENCED HYPERGLYCEMIA RELATED TO USE ERROR IN NOT PROPERLY SECURING THE BATTERY CAP TO THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
696309 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 69 YR Life Threatening