FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL0TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4212825
·
Received October 13, 2014
Report
- Report Number
- 2032227-2014-37310
- Event Type
- Malfunction
- Date Received
- October 13, 2014
- Date of Event
- August 30, 2014
- Report Date
- September 13, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INSULIN PUMP WAS RECEIVED WITH AN INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND MISSING ADDRESS AND SERIAL NUMBER LABEL.
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP HAS AN UNRESPONSIVE KEYPAD. CUSTOMER'S BLOOD GLUCOSE READING WAS 191 MG/DL. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 647450 | PARADIGM REAL0TIME REVEL INSULIN INFUSION PUMP | OYC | MEDTRONIC MINIMED | MMT-723LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |