FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4212802 · Received October 25, 2014

Report

Report Number
2032227-2014-43361
Event Type
Malfunction
Date Received
October 25, 2014
Date of Event
September 26, 2014
Report Date
September 26, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST AND PRIME TEST. HOWEVER, THE INSULIN PUMP WAS RECEIVED STUCK ON A MOTOR ERROR ALARM LOOP THAT OCCURRED DURING A BOLUS DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING PRIME. THE BLOOD GLUCOSE LEVEL WAS NOT INCLUDED IN THE REPORT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680662 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWS

Patients

Seq Age Sex Outcome Treatment
1