FDA Adverse Event Malfunction Summary report: N

NSK X85L

MDR report key: 4212537 · Received October 24, 2014

Report

Report Number
9611253-2014-00015
Event Type
Malfunction
Date Received
October 24, 2014
Date of Event
November 20, 2013
Report Date
October 23, 2014
Manufacturer
NAKANISHI, INC. (NSK)
Product Code
EGS
PMA / PMN Number
K972569
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

AS AN INVESTIGATIONAL APPROACH NAKANISHI INC., (B)(4) (MANUFACTURER) EXAMINED THE DHR FOR DEVICE (X85L, SERIAL NO. (B)(4)). NAKANISHI INC. CONCLUDED THAT NO PROBLEMS HAD OCCURRED DURING MANUFACTURING OR TESTING OF THE SUBJECT DEVICE, AS EVIDENCED IN THE DHR. (B)(6) CONCLUSION: NO RECORD OF REPAIR; NOISY UNIT MAY INDICATE A MAINTENANCE ISSUE. (B)(6) DID NOT RETURN HANDPIECE TO NSK FOR THE MANUFACTURER'S EVALUATION.

Description of Event or Problem · 1

THE FOLLOWING INFORMATION IS FROM (B)(6) TO NAKANISHI INC. (NSK), REGARDING A DEVICE MANUFACTURED BY NSK FOR (B)(6). (B)(6) EVENT SUMMARY: PATIENT BURNED DURING A CROWN PREPARATION #31. NICKEL SIZE BURN ON THE SIDE OF THE PATIENT'S TONGUE. NO PERMANENT SCARRING. DOCTOR NOTICED THE BURN, PATIENT NUMB. THE CUSTOMER NOTED THAT THE UNIT OVERHEATED AND BURNED (SEVERITY OF BURN NOT PROVIDED) A PATIENT'S TONGUE. THE ITEM WAS PURCHASED (B)(6) 2009, SO THE UNIT WAS ALMOST 53 MONTHS OLD AT THE TIME OF THE COMPLAINT. (B)(6) COMPLAINT REVIEW: THERE WAS NO REPAIR HISTORY ON FILE FOR THIS UNIT. (B)(6) INVESTIGATION: THE CUSTOMER NOTED THAT THE UNIT WOULD NOT BE REPAIRED, BUT WOULD BE TRADED IN TOWARDS A NEW HANDPIECE. THE (TRADE IN) UNIT WAS SENT IN THE REPAIR DEPARTMENT AND EVALUATED. THE REPAIR TECHNICIAN NOTED THAT THE UNIT DID NOT HEAT UP AT ALL WHEN CHECKED. THE TECHNICIAN DID FIND THAT THE CHUCK STRENGTH WAS POOR AND MAKING CLICKING NOISE. THE UNIT WAS REALLY LOUD AND THE COOL AND SPRAY WAS POOR AS WELL. THE UNIT WILL BE DISPOSED OF AND THE CUSTOMER PURCHASED A NEW HANDPIECE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
679562 NSK X85L HANDPIECE, CONTRA AND RIGHT-ANGLE ATTACHMENT, DENTAL EGS NAKANISHI, INC. (NSK) X85L BRA

Patients

Seq Age Sex Outcome Treatment
1 70 YR Other