FDA Adverse Event Malfunction Summary report: N

OT VERIO2 METER

MDR report key: 4212254 · Received October 30, 2014

Report

Report Number
2939301-2014-29077
Event Type
Malfunction
Date Received
October 30, 2014
Report Date
October 22, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K131363
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (12/26/2014). THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 11/13/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 12/12/2014 WITH THE FOLLOWING FINDINGS:THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 2 ¿ (02/06/2015). THE PATIENT¿S METER HAS BEEN RETURNED ON 12/2/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 1/22/2015 WITH THE FOLLOWING FINDINGS:THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE SPC PIN 3 LIFTED HIGH DUE TO A BAD CONTACT ON SPC PINS AND THE STRIP. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING APPLY SAMPLE - METER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
696170 OT VERIO2 METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3657413

Patients

Seq Age Sex Outcome Treatment
1