ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-30806
- Event Type
- Malfunction
- Date Received
- October 29, 2014
- Report Date
- October 15, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/08/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT UNEXPLAINABLE POWER REBOOTS WERE RECORDED. VISUAL INSPECTION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED FROM THE BUMPER PAD TO THE COVER. THE RETURNED BATTERY CAP WAS ABLE TO BE FULLY SECURED TO THE PUMP AND WAS USED TO COMPLETE ALL TESTING. THE BATTERY CAP CONTACT HEIGHT AND WIDTH MEASUREMENTS WERE FOUND TO BE WITHIN THE REQUIRED SPECIFICATIONS. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER INTERRUPTIONS OCCURRING. THE PUMP CASE WAS REMOVED AND EVIDENCE OF MOISTURE WAS FOUND ON THE PRINTED CIRCUIT BOARD AND OTHER INTERNAL COMPONENTS. THERE WAS NO DAMAGE TO THE POWER CIRCUIT OBSERVED. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE DISPLAY SCREEN WAS DISCOLORED. THE COMPLAINT THAT THE PUMP WAS LOSING POWER INTERMITTENTLY WAS OBSERVED IN THE PUMP HISTORY BUT WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THE REPORTER ALLEGED THAT THE PUMP WAS LOSING POWER INTERMITTENTLY. IT WAS NOTED THAT A NEW BATTERY WAS INSERTED INTO THE PUMP AND THE PUMP POWERED ON APPROPRIATELY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 691602 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR |