FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 4212127 · Received October 29, 2014

Report

Report Number
1416980-2014-37939
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
October 2, 2014
Report Date
October 2, 2014
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION, POWER ON SELF-TEST, BATTERY TESTING AND A REVIEW OF THE ALARM LOG WERE PERFORMED. DURING INVESTIGATION THE F-38 ALARM WAS FOUND; THE CAUSE OF THIS WAS INOPERATIVE FORCE SENSING RESISTORS (FSR). TO CORRECT THE CONDITION, THE FSRS WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT SERVICE BY A FIELD SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED AN F-38 ALARM. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693356 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1