FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4212060 · Received October 29, 2014

Report

Report Number
2032227-2014-44971
Event Type
Injury
Date Received
October 29, 2014
Date of Event
October 1, 2014
Report Date
October 1, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP POWERED UP PROPERLY AFTER BATTERY INSTALLATION. INSULIN PUMP RECEIVED WITH OPERATING CURRENTS WITHIN SPECIFICATIONS. NO FAILED BATTERY TEST ALARM NOTED. INSULIN PUMP PASSED REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO TRACES OF MOISTURE WERE NOTED AT ELECTRONIC OR MOTOR ASSEMBLIES PER VISUAL INSPECTION. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW AND RESERVOIR TUBE WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, BELT CLIP SLOT AND BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THEIR BUTTONS WERE STICKING ON THEIR INSULIN PUMP. THE CUSTOMER STATED, HE HAD EXPOSED HER INSULIN PUMP TO MOISTURE. THE CUSTOMER'S BLOOD GLUCOSE WAS 44 MG/DL. HE DECLINED TO TROUBLESHOOT FOR HIS BLOOD GLUCOSE. THE CUSTOMER ALSO STATED, HE ATTEMPTED TO ENTER HIS BLOOD GLUCOSE BUT HIS BUTTONS WERE UNRESPONSIVE. HE ALSO REPORTED RECEIVING A FAILED BATTERY TEST. THE CUSTOMER DECLINED TO TROUBLESHOOT FOR THE INSULIN PUMP. THE CUSTOMER WAS ADVISED TO DISCONNECT FORM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693249 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 26 YR