FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4212056
·
Received October 29, 2014
Report
- Report Number
- 2032227-2014-44965
- Event Type
- Malfunction
- Date Received
- October 29, 2014
- Date of Event
- October 1, 2014
- Report Date
- October 1, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
INSULIN PUMP HAD AN INTERMITTENT RESPONSE DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. NO SCROLLING NUMBERS DISPLAY ANOMALY NOTED. INSULIN PUMP RECEIVED WITH CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
IT WAS REPORTED THE NUMBERS WERE RAMPING ON THE CUSTOMER'S INSULIN PUMP. IT WAS ALSO REPORTED THE SCROLL BAR WAS MOVING UP OR DOWN WITHOUT INPUT FROM THE USER. CUSTOMER WAS ASKED TO RETURN THEIR INSULIN PUMP. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 694051 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522WWB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |