FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4212056 · Received October 29, 2014

Report

Report Number
2032227-2014-44965
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
October 1, 2014
Report Date
October 1, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD AN INTERMITTENT RESPONSE DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. NO SCROLLING NUMBERS DISPLAY ANOMALY NOTED. INSULIN PUMP RECEIVED WITH CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THE NUMBERS WERE RAMPING ON THE CUSTOMER'S INSULIN PUMP. IT WAS ALSO REPORTED THE SCROLL BAR WAS MOVING UP OR DOWN WITHOUT INPUT FROM THE USER. CUSTOMER WAS ASKED TO RETURN THEIR INSULIN PUMP. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
694051 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522WWB

Patients

Seq Age Sex Outcome Treatment
1