FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4212008 · Received October 29, 2014

Report

Report Number
2032227-2014-45048
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
October 2, 2014
Report Date
October 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH NO BUTTON RESPONSE, DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING NOTED. INSULIN PUMP RECEIVED WITH CRACKED CASE ON DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES AND CRACKED DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 198 MG/DL. THE CUSTOMER STATED SHE WAS ATTEMPTING TO GIVE HERSELF INSULIN PRIOR TO THE ALARM OCCURRING. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693406 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAL

Patients

Seq Age Sex Outcome Treatment
1 24 YR