FDA Adverse Event Malfunction Summary report: N

PRECISION XTRA

MDR report key: 4211957 · Received October 29, 2014

Report

Report Number
2954323-2014-01076
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
October 9, 2014
Report Date
October 9, 2014
Manufacturer
ABBOTT DIABETES CARE INC.
Product Code
NBW
PMA / PMN Number
K040814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REQUESTED PRODUCT WAS NOT RECEIVED FOR INVESTIGATION. RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER (B)(4) WERE TESTED WITH CONTROL SOLUTION INSTEAD. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 16 MMOL/L AND 3 MMOL/L WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693853 PRECISION XTRA BLOOD GLUCOSE MONITORING SYSTEM NBW ABBOTT DIABETES CARE INC. 4500163231

Patients

Seq Age Sex Outcome Treatment
1