PRECISION XTRA
Report
- Report Number
- 2954323-2014-01076
- Event Type
- Malfunction
- Date Received
- October 29, 2014
- Date of Event
- October 9, 2014
- Report Date
- October 9, 2014
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- NBW
- PMA / PMN Number
- K040814
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
REQUESTED PRODUCT WAS NOT RECEIVED FOR INVESTIGATION. RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER (B)(4) WERE TESTED WITH CONTROL SOLUTION INSTEAD. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED.
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 16 MMOL/L AND 3 MMOL/L WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 693853 | PRECISION XTRA | BLOOD GLUCOSE MONITORING SYSTEM | NBW | ABBOTT DIABETES CARE INC. | 4500163231 |