FDA Adverse Event Malfunction Summary report: N

SPECTRANETICS EXCIMER LASER SYSTEM

MDR report key: 4211931 · Received October 29, 2014

Report

Report Number
1721279-2014-00192
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
October 16, 2014
Report Date
October 16, 2014
Manufacturer
SPECTRANETICS CORPORATION
Product Code
LPC
PMA / PMN Number
P910001
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EXCIMER LASER EVALUATION (PERFORMED AT THE FACILITY): UPON ARRIVAL, FSE CONFIRMED FAULT CODE 04 DURING CALIBRATION. FOUND HVPS LOW AND REGULATED ENERY LOW. ADJUSTED HVPS AND MONITORS. SYSTEM STABLE AFTER 10 MINUTE RUN. COMPLETED CAL FORM. SYSTEM FUNCTIONING TO SPECIFICATION.

Description of Event or Problem · 1

THIS WAS A LEAD EXTRACTION CASE PERFORMED IN THE HYBRID OR TO REMOVE TWO NON-FUNCTIONING LEADS FROM THE RA (MDT 5568 AND 5076, IMPLANTED FOR 89 MONTHS). THE FIRST LEAD (MDT 5568) WAS SUCCESSFULLY EXTRACTED WITH NO ISSUES. THE SECOND LEAD (MDT 5076) WAS PREPPED WITH AN LLD-EZ AND A LASER CATHETER WAS USED FOR THE EXTRACTION. DURING THE PROCEDURE, THE LASER GAVE A FAULT CODE, AND STOPPED WORKING. THE MANUFACTURER ENGINEER WAS CALLED AND INSTRUCTIONS WERE PROVIDED TO TRY AND RESET THE LASER. AFTER TRYING ALL POSSIBLE REMEDIES, AND THE LASER NOT FUNCTIONING, THE PHYSICIAN DETERMINED THE BEST METHOD TO TREAT THE PATIENT WAS TO LEAVE THE REMAINING LEAD IN PLACE. THERE WERE NO ALTERNATIVE MEANS AVAILABLE FOR THIS PHYSICIAN TO EXTRACT THE LEAD. THE PATIENT HAD AN OCCLUDED SVC AND THE PHYSICIAN WAS UNABLE TO PUT IN A NEW RA LEAD. THE PHYSICIAN CUT AND CAPPED THE 5076 LEAD, LEAVING A PART OF THE LLD IN THE LEAD IN THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692938 SPECTRANETICS EXCIMER LASER SYSTEM CVX-300 EXCIMER LASER LPC SPECTRANETICS CORPORATION CVX-300 NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR Other LEAD LOCKING DEVICE| VISISHEATH DILATOR SHEATH| GLIDELIGHT LASER SHEATH