FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4211905 · Received October 29, 2014

Report

Report Number
2032227-2014-45017
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
October 1, 2014
Report Date
October 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER BLOOD GLUCOSE WAS NOT PROVIDED. CUSTOMER'S MOTHER STATED THE DEVICE ALSO ALARMED LOW SUSPEND. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692910 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-754LWWS

Patients

Seq Age Sex Outcome Treatment
1