FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4211667 · Received October 29, 2014

Report

Report Number
1416980-2014-37922
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
October 4, 2014
Report Date
October 4, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT NO: CMPLNT-(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, IMPROPERLY DISCONNECTING/RECONNECTING FROM THE SET IS A KNOWN CAUSE OF THIS ALARM. PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE HAS INSTRUCTIONS ON THE STEPS TO PROPERLY DISCONNECT FROM CYCLER DURING AN EMERGENCY AND EXPLAINS HOW TO RECONNECT FOR THERAPY AFTER PROPERLY DISCONNECTING. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2240 ALARM (AIR IN SET / LINE) OCCURRED ON THE HOMECHOICE (HC) DEVICE DURING USE. DURING THE TROUBLESHOOTING, IT WAS DETERMINED THAT THE HOME PATIENT HAD DISCONNECTED FROM THE SET AND RECONNECTED BACK BUT DID NOT FOLLOW PROPER DISCONNECT PROCEDURES. THE TECHNICAL SERVICE REPRESENTATIVE REVIEWED PROPER DISCONNECT PROCEDURE AND HAD THE CARE GIVER (CG) CYCLE THE POWER ON THE HC TO CLEAR THE ALARM. THE CG WOULD START OVER WITH NEW SUPPLIES TO COMPLETE THE THERAPY FOR THE HP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692522 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 79 YR HOMECHOICE