FDA Adverse Event Malfunction Summary report: N

DA VINCI SI SURGICAL SYSTEM

MDR report key: 4211651 · Received October 29, 2014

Report

Report Number
2955842-2014-05569
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
September 19, 2014
Report Date
September 30, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K081137
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT SIDE MANIPULATOR (PSM) ARM 2 WAS RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING FOUND THAT THE PSM ARM 2 PASSED SINE CYCLE AND SLOW SWEEP, BUT FAILED THE IN-HOUSE BRAKE DROP TEST. THE AXIS-2 BRAKE, AXIS-2 GEARS, BRAKE BEARINGS AND BRAKE SHAFT WERE REPLACED. THIS COMPLAINT IS BEING REPORTED DUE TO THE PATIENT SIDE MANIPULATOR ARM FAILING THE BRAKE WEIGHT DROP TEST DURING FAILURE ANALYSIS. ALTHOUGH THIS WAS FOUND DURING FAILURE ANALYSIS AND NO PATIENT INVOLVEMENT OCCURRED, IF THIS MALFUNCTION WERE TO RECUR IT COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING MULTIPLE DA VINCI SURGICAL PROCEDURES, THE SURGEON NOTED MOVEMENT OF INSTRUMENTS IN ARM 2 WERE STICKING AND JERKING. ARM 3 HAD NOT BEEN USED AND WAS CURRENTLY DOCKED. THE SURGEON SWITCHED TO PATIENT SIDE MANIPULATOR (PSM) ARM 3 CONTROLED BY MASTER TOOL MANIPULATOR (MTM) AND IT FELT NORMAL. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692128 DA VINCI SI SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS3000 A60P8

Patients

Seq Age Sex Outcome Treatment
1