PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-44909
- Event Type
- Malfunction
- Date Received
- October 29, 2014
- Date of Event
- October 1, 2014
- Report Date
- October 1, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
NO UNEXPECTED BUTTON ERROR ALARM WAS NOTED. NO UNEXPECTED FAILED BATTERY TEST ALARM WAS NOTED. THE INSULIN PUMP PASSED THE DISPLACEMENT TEST AND THE OFF NO POWER TEST. THE OPERATING CURRENTS WERE WITHIN SPECIFICATION. THE INSULIN PUMP HAD INTERMITTENT ACT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE WINDOW, CRACKED RESERVOIR TUBE, CRACKED BATTERY TUBE THREADS, A CRACKED BELT CLIP SLOT AND A STAINED END CAP STICKER.
CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP DURING A BOLUS. CUSTOMER ALSO RECEIVED A FAILED BATTERY TEST. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 211 MG/DL. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 692480 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |