FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4211633
·
Received October 29, 2014
Report
- Report Number
- 2032227-2014-44907
- Event Type
- Malfunction
- Date Received
- October 29, 2014
- Date of Event
- October 2, 2014
- Report Date
- October 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).UNIT UNABLE TO PRIME DURING THE PRIME/A33 TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS CRACKED, BROKEN BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP, RESERVOIR TUBE CRACKED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP HAD DAMAGE NEAR THE RESERVOIR AND THE RIGHT AMOUNT OF INSULIN WAS NOT COMING OUT. CUSTOMER HAS HAD TO TREAT WITH MANUAL INJECTIONS TO BRING THEIR BLOOD GLUCOSE READINGS DOWN. CUSTOMER MENTIONED THAT A PIECE OF THE INSULIN PUMP WAS MISSING. CUSTOMER'S MOST RECENT BLOOD GLUCOSE READING WAS 384 MG/DL AND HAS TREATED WITH THE INSULIN PUMP. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 692111 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |