CD001, 10MM RETRIEVAL SYSTEM, 10/BX
Report
- Report Number
- 2027111-2014-00363
- Event Type
- Malfunction
- Date Received
- October 29, 2014
- Date of Event
- September 16, 2014
- Report Date
- December 1, 2014
- Manufacturer
- APPLIED MEDICAL
- Product Code
- GCJ
- PMA / PMN Number
- K060051
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
RA HAS JUST RECEIVED THE INCIDENT DEVICE AND HAS BEEN ASSIGNED TO ENGINEERING FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SENT UPON COMPLETION OF INVESTIGATION. IN ACCORDANCE TO 21 CFR 803.56, IF WE OBTAIN ADDITIONAL INFORMATION, WHICH WAS NOT KNOWN OR WAS NOT AVAILABLE WHEN THE INITIAL REPORT WAS SUBMITTED, THEN THE SUPPLEMENTAL REPORT WILL BE SUBMITTED TO THE FDA.
INVESTIGATION SUMMARY: UPON INSPECTION, ENGINEERING CONFIRMED A SMALL TEAR NEAR THE BOTTOM OF THE BAG. THE BAG SHOWED SIGNS OF TENSION, CLEARLY VISIBLE IN THE STRETCHING NEAR THE BURST SITE. PREVIOUS TESTS HAVE SHOWN IDENTICAL STRETCHING AND BREAKING IN RETRIEVAL BAGS WHEN EXCESSIVE FORCE AND MANIPULATION IS USED TO REMOVE SPECIMENS FROM A SMALL INCISION; HOWEVER, FRAGMENTATION OF THE BAG HAS NOT BEEN OBSERVED DURING TESTING. THE INSTRUCTIONS FOR USE STATE, "NOTE: IF THE BAG AND ITS CONTENTS ARE TOO LARGE TO BE EXTRACTED, CAREFULLY ENLARGE THE PORT SITE FOR EASE OF BAG REMOVAL." WHEN THE INCISION IS ENLARGED, THE SPECIMEN CAN BE REMOVED WITHOUT INCIDENT. THIS DOCUMENT REPRESENTS OUR FINAL REPORT.
LAP CHOLE- "ENDOSCOPIC BAG RUPTURED. GALL BLADDER AND CONTENTS INSIDE WERE SPILLED IN PATIENT'S ABDOMEN. CONTENTS REMOVED AND ABD WAS IRRIGATED UNTIL CLEAR IRRIGATION, WAS EVACUATED. THIRD BAG NOTED TO LET GO WITH MANUAL FORCE." ((B)(6)).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 692068 | CD001, 10MM RETRIEVAL SYSTEM, 10/BX | GCJ | GCJ | APPLIED MEDICAL | CD001 | 1221801 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |