SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2014-37910
- Event Type
- Malfunction
- Date Received
- October 29, 2014
- Date of Event
- October 4, 2014
- Report Date
- October 4, 2014
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, IT WAS REPORTED THAT THERE WAS A LOOSE CONNECTION WHICH IS A KNOWN CAUSE OF THIS ALARM. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM OCCURRED DURING FILL ONE ON THE HOMECHOICE. THE HOME PATIENT WAS CONNECTED AT THE TIME OF THE EVENT. THE CAREGIVER STATED THE HEATER BAG HAD A LOOSE CONNECTION WITH THE CASSETTE LINE. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED THE CAREGIVER TO END THERAPY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED AS A RESULT OF THIS EVENT. N ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 692546 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HOMECHOICE |