FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4211100 · Received October 29, 2014

Report

Report Number
1416980-2014-37910
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
October 4, 2014
Report Date
October 4, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED.  AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, IT WAS REPORTED THAT THERE WAS A LOOSE CONNECTION WHICH IS A KNOWN CAUSE OF THIS ALARM. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM OCCURRED DURING FILL ONE ON THE HOMECHOICE. THE HOME PATIENT WAS CONNECTED AT THE TIME OF THE EVENT. THE CAREGIVER STATED THE HEATER BAG HAD A LOOSE CONNECTION WITH THE CASSETTE LINE. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED THE CAREGIVER TO END THERAPY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED AS A RESULT OF THIS EVENT. N ADDITIONAL INFORMATION IS AVAILABLE.