ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-30796
- Event Type
- Malfunction
- Date Received
- October 29, 2014
- Report Date
- October 17, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/20/2014 WITH THE FOLLOWING FINDINGS: ON (B)(6) 2014, THE PUMP'S BLACK BOX RECORDED THREE ¿LOSS OF PRIME¿ WARNINGS DUE TO A LOW, NON-ZERO, FORCE SENSOR READING. PUMP WAS EXERCISED FOR 24 HOURS AT 1 UNIT/HOUR WITH NO ¿LOSS OF PRIME¿ WARNING OR OTHER ALARMS OBSERVED. THE PUMP'S FORCE SENSOR WAS FOUND TO BE CALIBRATED WITHIN SPECIFICATION. THE ALLEGED ISSUE COULD NOT BE DUPLICATED DURING TESTING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. IT WAS REPORTED THAT A LOSS OF PRIME WARNING OCCURRED 3 OR MORE TIMES WITHIN A 30 DAY PERIOD, DURING WHICH AT LEAST 3 SEPARATE CARTRIDGES FROM 2 SEPARATE BOXES WERE USED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 691604 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |