FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4210937 · Received October 29, 2014

Report

Report Number
2032227-2014-45031
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
October 1, 2014
Report Date
October 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO UNEXPECTED BUTTON ERROR ALARMS NOTED. INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE ON THE ACT BUTTON DUE TO CORRODED KEYPAD TRACES NOTED. INSULIN PUMP HAD A CRACKED LCD WINDOW, CRACKED CASE AT LCD WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, SCRATCHES ON RESERVOIR TUBE WINDOW, CRACKED BATTERY TUBE THREADS AND CRACKED BELT CLIP SLOT.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ON HER INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE IS 70 MG/DL. THERE WERE NO SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER WAS ADVISED THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693329 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAB

Patients

Seq Age Sex Outcome Treatment
1 19 YR