PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-44762
- Event Type
- Malfunction
- Date Received
- October 29, 2014
- Date of Event
- September 30, 2014
- Report Date
- September 30, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- OTHER
Narratives
(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE PUMP WAS RECEIVED WITH A MOTOR ERROR ALARM DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL. THE PUMP ALSO HAD A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED DISPLAY WINDOW, AND A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS.
CUSTOMER REPORTED RECEIVING A FAILED BATTERY TEST ALARM AND A MOTOR ERROR DURING PRIMING. CUSTOMER DOES NOT USE THE SENSOR FEATURE AND THEY WERE UNABLE TO COMPLETE THE REWIND SEQUENCE. CUSTOMER'S BLOOD GLUCOSE READINGS AT THE TIME OF THE CALL WERE 237 MG/DL. NO FURTHER INFORMATION REPORTED.