FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4210909 · Received October 29, 2014

Report

Report Number
2032227-2014-44762
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
September 30, 2014
Report Date
September 30, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE PUMP WAS RECEIVED WITH A MOTOR ERROR ALARM DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL. THE PUMP ALSO HAD A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED DISPLAY WINDOW, AND A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A FAILED BATTERY TEST ALARM AND A MOTOR ERROR DURING PRIMING. CUSTOMER DOES NOT USE THE SENSOR FEATURE AND THEY WERE UNABLE TO COMPLETE THE REWIND SEQUENCE. CUSTOMER'S BLOOD GLUCOSE READINGS AT THE TIME OF THE CALL WERE 237 MG/DL. NO FURTHER INFORMATION REPORTED.