FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4210877 · Received October 29, 2014

Report

Report Number
2032227-2014-44758
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
September 30, 2014
Report Date
September 30, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, A CRACKED CASE AT THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A SCRATCHED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED NO DELIVERY AND MOTOR ERROR DURING BOLUS. CUSTOMER DOES NOT USE THE SENSOR FEATURE AND THEY WERE ABLE TO COMPLETE THE REWIND SEQUENCE. CUSTOMER DECLINED ANY FURTHER TROUBLESHOOTING. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 53 MG/DL AND HAS TREATED WITH FOOD. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693253 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 7 YR