FDA Adverse Event Injury Summary report: N

ELLIPSE ST VR

MDR report key: 4210687 · Received October 29, 2014

Report

Report Number
2938836-2014-17204
Event Type
Injury
Date Received
October 29, 2014
Date of Event
August 20, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
LWS
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.THE REPORTED FIELD EVENT OF BACKUP VVI WAS CONFIRMED IN THE LABORATORY AND WAS CAUSED BY A POWER-ON RESET. THE DEVICE WAS TESTED ON THE BENCH AS WELL AS USING AUTOMATED TEST EQUIPMENT AND THE HV OUTPUT CIRCUIT WAS FOUND TO BE DAMAGED. THE CAUSE OF THE DAMAGED HV OUTPUT CIRCUIT AND POWER-ON RESET COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD A VT ABLATION. THE PHYSICIAN REPORTED THAT THE DEVICE DELIVERED THERAPY SUCCESSFULLY DURING THE ABLATION. THREE HOURS POST ABLATION, THE PATIENT WENT INTO A VF ELECTRICAL STORM. THE DEVICE SUCCESSFULLY SHOCKED AND TERMINATED THE VF, BUT SEVERAL MINUTES LATER, THE VF STORM CONTINUED AND SHOCKS WERE NOT SUCCESSFUL. THE RHYTHM WAS TERMINATED WITH EXTERNAL DEFIBRILLATION. THE NEXT DAY, DEVICE INTERROGATION REVEALED BACK-UP VVI MODE. IT WAS NOT KNOWN WHAT CAUSED THE BACK-UP VVI MODE. THE DEVICE WAS EXPLANTED AND REPLACED. THE PATIENT CONDITION WAS GOOD AFTER THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693668 ELLIPSE ST VR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST. JUDE MEDICAL, INC., CRMD CD1277-36 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention (B)(4)