FDA Adverse Event
Injury
Summary report: N
QUADRA ASSURA CRT-D, DF-4 CONNECTOR
MDR report key: 4210654
·
Received October 29, 2014
Report
- Report Number
- 2938836-2014-17185
- Event Type
- Injury
- Date Received
- October 29, 2014
- Date of Event
- September 25, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NIK
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE SYSTEM WAS EXPLANTED DUE TO INFECTION. PATIENTS CONDITION WAS STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 693861 | QUADRA ASSURA CRT-D, DF-4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC., CRMD | CD3365-40Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Required Intervention | (B)(4) |