FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4210582 · Received October 25, 2014

Report

Report Number
2032227-2014-43524
Event Type
Injury
Date Received
October 25, 2014
Date of Event
July 1, 2014
Report Date
September 26, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE UNIT PASSED FUNCTIONAL TEST INCLUDING DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33, AND EXCESSIVE NO DELIVERY ALARM TEST. THE UNIT FUNCTIONED PROPERLY. THE UNIT WAS RECEIVED WITH A CRACKED CASE. THE CRACKS WERE ON THE DISPLAY WINDOW CORNERS, MINOR SCRATCHES ON THE LCD WINDOW, A CRACKED RESERVOIR TUBE LIP, AND CRACKED BELT CLIP SLOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED ON (B)(6) 2014 DUE TO A HIGH BLOOD GLUCOSE LEVEL OF 600 MG/DL. SHE STAYED OVERNIGHT FOR TREATMENT. THE CUSTOMER ALSO REPORTED A RECENT HIGH BLOOD GLUCOSE LEVEL OF 500 MG/DL. HER BLOOD GLUCOSE LEVEL AT THE TIME OF THE REPORT WAS 76 MG/DL. SHE HAD BEEN EXPERIENCING LARGE FLUCTUATIONS IN HER BLOOD GLUCOSE LEVELS. SHE HAD BEEN USING MANUAL INJECTIONS TO TREAT HER HIGH BLOOD GLUCOSE LEVELS. STANDARD TROUBLESHOOTING WAS PERFORMED. CRACKS IN THE PUMP WERE DISCOVERED. THE PRODUCT HAS BEEN RETURNED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680645 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC MEDTRONIC MINIMED MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 57 YR Hospitalization