LEVEEN? COACCESS?

MDR report key: 4210569 · Received October 29, 2014

Report

Report Number
2134265-2014-07009
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
September 30, 2014
Report Date
October 1, 2014
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
GEI
PMA / PMN Number
K012315
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER - RECEIVED ONE COACCESS ELECTRODE, CABLE AND INTRODUCER WITH NO VISIBLE RESIDUE FOUND ON THE DEVICE. THE INTRODUCER AND CABLE WERE WITHOUT ISSUE. A VISUAL EXAMINATION OF THE DEVICE FOUND THE TINES TO BE FULLY RETRACTED. A FUNCTIONAL EVALUATION WAS PERFORMED BY EXTENDING AND RETRACTING THE ARRAY WITHOUT ISSUE. THE TINES WERE FOUND TO BE UN-EVENLY SPACED AND BENT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DIFFICULTY EXTENDING THE NEEDLE OCCURRED. WHEN FUNCTIONAL TESTING WAS PERFORMED DURING PREPARATION, THE TARGET SITE WAS PUNCTURED WITH THE LEVEEN¿ COACCESS¿ ELECTRODE SYSTEM 4.0/15/15, HOWEVER, THE NEEDLE FAILED TO FULLY EXTEND. THE PROCEDURE WAS COMPLETED WITH ANOTHER LEVEEN¿ COACCESS¿ SYSTEM. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT¿S STATUS WAS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DIFFICULTY EXTENDING THE NEEDLE OCCURRED. WHEN FUNCTIONAL TESTING WAS PERFORMED DURING PREPARATION, THE TARGET SITE WAS PUNCTURED WITH THE LEVEEN COACCESS ELECTRODE SYSTEM 4.0/15/15 HOWEVER THE NEEDLE FAILED TO FULLY EXTEND. THE PROCEDURE WAS COMPLETED WITH ANOTHER LEVEEN COACCESS SYSTEM. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
691962 LEVEEN? COACCESS? ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI BOSTON SCIENTIFIC - SPENCER M001262240 16624973