ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Report
- Report Number
- 3002648230-2014-00182
- Event Type
- Injury
- Date Received
- October 29, 2014
- Date of Event
- October 1, 2014
- Report Date
- October 1, 2014
- Manufacturer
- MEDTRONIC CRYOCATH LP
- Product Code
- OAE
- PMA / PMN Number
- P100010/S015
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE IS NOT AVAILABLE FOR INVESTIGATION; IT WAS DISCARDED AFTER THE PROCEDURE. BIN FILES WERE REVIEWED AND DID NOT SHOW ANY SYSTEM NOTICE MESSAGES FOR THE DATE OF EVENT. BIN FILES SHOWED AT LEAST 15 INJECTIONS WERE PERFORMED WITH THE CATHETER. THERE WAS NO INDICATION OF PRODUCT MALFUNCTION.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED INFORMATION OF A PHRENIC NERVE PARALYSIS THAT OCCURRED AT THE CRYOABLATION OF THE RIGHT SUPERIOR PULMONARY VEIN ( RSPV). CRYOABLATION OF RSPV WAS PERFORMED AFTER THE CRYOABLATION OF THE LEFT PULMONARY VEINS AND THE RIGHT INFERIOR PULMONARY VEIN. EIGHTY SECONDS AFTER THE START OF RSPV CRYOABLATION, THE DIAPHRAGM RESPONSE ON PALPATION WAS DECREASED AND THE ABLATION WAS STOPPED. THE CRYOABLATION PROCEDURE WAS COMPLETED AND THE PATIENT IS BEING MONITORED. A CHEST X-RAY PERFORMED ON (B)(6) 2014 CONFIRMED RECOVERY OF THE PHRENIC NERVE PARALYSIS.
MEDTRONIC RECEIVED INFORMATION OF A PHRENIC NERVE PARALYSIS THAT OCCURRED AT THE CRYOABLATION OF THE RIGHT SUPERIOR PULMONARY VEIN (RSPV). CRYOABLATION OF RSPV WAS PERFORMED AFTER THE CRYOABLATION OF THE LEFT PULMONARY VEINS AND THE RIGHT INFERIORPULMONARY VEIN. 80 SECONDS AFTER THE START OF RSPV CRYOABLATION, THE DIAPHRAGM RESPONSE ON PALPATION WAS DECREASED AND THE ABLATION WAS STOPPED. THE CRYOABLATION PROCEDURE WAS COMPLETED AND THE PATIENT IS BEING MONITORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 691374 | ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER | PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION | OAE | MEDTRONIC CRYOCATH LP | 2AF284 | 11940 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00057 YR | Other |