FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC HEAT PACK

MDR report key: 4210349 · Received August 8, 2014

Report

Report Number
1811605-2014-00169
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
July 16, 2014
Report Date
August 8, 2014
Manufacturer
BATTLE CREEK EQUIPMENT
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

OUR INSPECTION FOUND A SMALL CUT IN THE CORD NEAR THE SWITCH THAT COULD HAVE EXPOSED THE USER TO BARE WIRE. THE INSPECTION ALSO REVEALED CUSTOMER MISUSE AS OUR INSPECTION FOUND: PAD BUNCHED, BENT/BROKEN LEAD, THERMOSTAT DISCOLORATION, PAD DIRTY OR STAINED, PAD STITCHING LOOSE/OPEN. THE PAD ALSO APPEARED TO HAVE BEEN LAYED ON. THIS TELLS US THAT THE PAD WAS NOT BEING PROPERLY USED AS PER OUR INSTRUCTIONS IN THE MANUAL. CORD BREAKAGE USUALLY OCCURS WHEN THE CORD IS REPEATEDLY OVER BENT NEAR THE SWITCH. WE HAVE INITIATED A CAPA PROJECT (B)(4) TO REDUCE THE OCCURRENCE OF THIS ISSUE. A NEW JACKETED CORD DESIGN WAS LAUNCHED IN PRODUCTION ON 03/10/2014 THAT CLOSES (B)(4) PENDING A FUTURE EVALUATION OF THE EFFECTIVENESS OF THE SOLUTION.

Description of Event or Problem · 1

CUSTOMER CALLED AND STATED THAT CORD STARTED SMOKING, AND IT IS FRAYED BY THE TRIGGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470558 THERMOPHORE AUTOMATIC HEAT PACK ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT 055