PARADIGM REAL -TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-37535
- Event Type
- Malfunction
- Date Received
- October 13, 2014
- Date of Event
- September 12, 2014
- Report Date
- September 13, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP MAY NOT BE FUNCTIONING PROPERLY DUE TO HIGH BLOOD GLUCOSE LEVELS. CUSTOMER STATED THAT THEY HAD HIGH BLOOD GLUCOSE LEVEL OF OVER 500 MG/DL AND THAT THE DEVICE MAY NOT BE DELIVERING INSULIN. CUSTOMER ADVISED TO DISCONNECT FROM PUMP. DURING TROUBLESHOOTING, CUSTOMER WAS ASKED TO CHECK THE CONDITION OF THE DRIVE SUPPORT CAP ON THE DEVICE AND FOUND IT TO BE RECESSED. NEXT, CUSTOMER WAS ASKED TO CHECK FOR AIR BUBBLES IN INFUSION SET TUBING. CUSTOMER VERIFIED THERE WERE NO AIR BUBBLES. AFTERWARDS, THE RESERVOIR WAS REINSERTED INTO INSULIN PUMP AND A MANUAL PRIME WAS PERFORMED. DEVICE PASSED PRIMING, HIGH PRESSURE TEST AND INSULIN EXITED TUBING. CUSTOMER VERIFIED DEVICE WAS NOT LEAKING AND DID NOT ALARM. INFORMED CUSTOMER THE DEVICE IS FUNCTIONING PROPERLY. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 647247 | PARADIGM REAL -TIME REVEL INSULIN INFUSION PUMP | OYC | MEDTRONIC MINIMED | MMT-723RNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |