FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4210339 · Received October 29, 2014

Report

Report Number
2032227-2014-44709
Event Type
Malfunction
Date Received
October 29, 2014
Date of Event
September 30, 2014
Report Date
September 30, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE REWIND, DISPLACEMENT TEST, PRIME TEST, AND EXCESSIVE NO DELIVERY ALARM TEST. NO EXCESSIVE NO DELIVERY ALARM WAS NOTED. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE HAS RECEIVED SOME NO DELIVERY ALARMS. CUSTOMER HAS TRIED DIFFERENT INFUSION SETS WITH DIFFERENT CANNULA LENGTHS AND HAS NOT RESOLVED THE ISSUE. CUSTOMER STATED THAT THE TUBING IS NOT BENT OR KINKED. CUSTOMER STATED THAT THE PISTON ON THE INSULIN PUMP IS GETTING STUCK. BLOOD GLUCOSE VALUE WAS 278 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
691981 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 30 YR