ADVIA CHEMISTRY HEMOGLOBIN A1C_3 CALIBRATOR
Report
- Report Number
- 2432235-2014-00617
- Event Type
- Malfunction
- Date Received
- October 28, 2014
- Date of Event
- September 10, 2014
- Report Date
- May 3, 2017
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JIX
- PMA / PMN Number
- K081895
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A SIEMENS HEADQUARTERS SUPPORT CENTER (HSC) REPRESENTATIVE HAS EVALUATED THE PATIENT DATA. THE CAUSE OF THE INITIAL LOW RESULT FOR A1C3 WHEN USING CALIBRATOR LOT 3DD010 IS UNKNOWN. SIEMENS IS INVESTIGATING THE ISSUE.
THE INITIAL MDR 2432235-2014-00617 WAS FILED ON OCTOBER 28, 2014. ADDITIONAL INFORMATION (2/12/2015): SIEMENS HEADQUARTERS SUPPORT CENTER (HSC) EVALUATED THE DATA PROVIDED BY THE CUSTOMER. HSC IDENTIFIED INCIDENCES OF CLOTTED SAMPLES BEING RUN ON THE INSTRUMENT. THE CUSTOMER WAS NOT FOLLOWING MANUFACTURER'S INSTRUCTIONS ON SAMPLE HANDLING. THE CAUSE OF THE INITIAL HIGH RESULTS FOR A1C_3 IS UNKNOWN. THIS INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.
THE CUSTOMER HAS OBTAINED A LOWER VALUE FOR A1C3 ON ONE PATIENT SAMPLE ON AN ADVIA 2400 INSTRUMENT WHEN USING CALIBRATOR LOT 3DD010. THE SAMPLE WAS TESTED ON SEPTEMBER 10, 2014 FOR A1C3 ON THE ADVIA 2400. THE PATIENT WENT TO ANOTHER FACILITY AND GOT THEMSELVES TESTED ON AN ALTERNATE PLATFORM AND THE RESULT WAS HIGHER. THREE DAYS AFTER THE EVALUATION THE PATIENT HAD TO BE HOSPITALIZED DUE TO ACUTE SIGNS OF HYPERGLYCEMIA. IT IS UNKNOWN IF THE INITIAL RESULT ON THE ADVIA 2400 OR THE RESULT ON THE ALTERNATE PLATFORM WERE REPORTED TO PHYSICIAN(S). THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO INITIAL LOW RESULT FOR A1C3 ON THE ONE PATIENT SAMPLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 689993 | ADVIA CHEMISTRY HEMOGLOBIN A1C_3 CALIBRATOR | ADVIA CHEMISTRY HEMOGLOBIN A1C_3 CALIBRATOR | JIX | SIEMENS HEALTHCARE DIAGNOSTICS INC. | ADVIA CHEMISTRY A1C_3 CALIBRATOR | 3DD010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |