FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4205490
·
Received October 28, 2014
Report
- Report Number
- 1031452-2014-16399
- Event Type
- Malfunction
- Date Received
- October 28, 2014
- Report Date
- October 7, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
THE DEALER CALLED IN TO TROUBLESHOOT WITH TECH HE WAS GETTING ONE RED TWO 2 GREEN LIGHTS, HIGH PRESSURE NOT SWITCHING. UNIT SHUTTING DOWN WITH 1R, 2GR CODE. DEALER HAS REPLACED BEDS, 4 WAY AND HAS OHM'S OUT PILOT VALE AT 180 OHMS. SOUNDS LIKE PILOT VALVE COULD BE HANGING UP. SENDING PCB ALSO IF PILOT VALVE DOESN'T FIX ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 688284 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |