FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4205490 · Received October 28, 2014

Report

Report Number
1031452-2014-16399
Event Type
Malfunction
Date Received
October 28, 2014
Report Date
October 7, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

THE DEALER CALLED IN TO TROUBLESHOOT WITH TECH HE WAS GETTING ONE RED TWO 2 GREEN LIGHTS, HIGH PRESSURE NOT SWITCHING. UNIT SHUTTING DOWN WITH 1R, 2GR CODE. DEALER HAS REPLACED BEDS, 4 WAY AND HAS OHM'S OUT PILOT VALE AT 180 OHMS. SOUNDS LIKE PILOT VALVE COULD BE HANGING UP. SENDING PCB ALSO IF PILOT VALVE DOESN'T FIX ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
688284 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other