FDA Adverse Event Malfunction Summary report: N

CLINITRON BED

MDR report key: 4203554 · Received September 18, 2014

Report

Report Number
1824206-2014-02287
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
August 21, 2014
Report Date
August 21, 2014
Manufacturer
HILL-ROM, INC.
Product Code
INX
PMA / PMN Number
K914351
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BRAKES WOULD SET BUT NOT HOLD. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THIS BEDS. TECHNICIAN REPLACED BOTH BRAKE CATERS TO RESOLVE THE ISSUE. BASED. ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES ARE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
578790 CLINITRON BED BED, AIR FLUIDIZED INX HILL-ROM, INC. 0800

Patients

Seq Age Sex Outcome Treatment
1